Laboratory Glove Box

Description

Model Number:Zoway-2working
1. Compliance:CE Certified
2. Warranty:2 Year limited warranty with lifetime support
3. Delivery Time:In Stock

Email: imani@batterypackassembly.com
Tel:+86 13174506016
WhatsApp:+86 13174506016
WeChat:18659217588

Content

Email: imani@batterypackassembly.com

FEATURES

ZOWAY-2working is a dual stainless steel glove box designed for researchers in material science, chemistry, semiconductor, and related topics. The purification cylinder is made of BASF antioxidant materials from Germany and UOP water adsorbent materials from the US in order to achieve a low water and oxygen concentration of lower than 1ppm, such that experiments can be conducted in an ultra clean and highly purified gas environment. It is integrated with vapor pressure control system, purity control system, automatic gas purity regeneration control system, auto alarm, and system data log recording. TMAX-2working is widely used in preparing nanomaterials, Li-ion battery materials, catalysts, and metallorganics. 


SPECIFICATIONS:

Working Voltage

AC 110 - 240 V 50/60 Hz

Max. Power Consumption

3KW

Glove Box Chamber

Dimensions: 2440mm(L) x 750mm(W) x 900mm(H)
Case Material: Stainless Steel 304
12mm thick double layer laminated safety glass with corrosion protection film for operating in a safe and easy manner.
<0.05 Vol% / hr (Depending on oxygen concentration inside the glove box)
Built-in fluorescent lamp
Two slidable shelves, 300mm L x 250mm W x 300mm H

Airlock Chamber

Antechamber Dimensions: 360mm(Dia.) x 600mm(L)
Small Chamber Dimensions (for tools): 150mm(Dia.) x 300mm(L)
Max. vacuum level: 100 Pa

Two Transparent Front Panels

    1.Front Panel made of transparent plastic 
      2.Each Panel Dimension: 1170 mm(L) x 700mm(W)
        3.Removable for putting in devices

                Attainable purity level

                  1.Water concentration: <1 ppm (20°C, 1 atm)
                    2.Oxygen concentration: <1 ppm (20°C, 1 atm)

                      Working Gas

                        1.Operating gas: Inert gas such as N2, Ar, He
                          2.Regenerating gas: Mixture of H2 (5%) and Operating gas

                            Gas Purification System

                            Auto regeneration process
                            Auto purging function
                            Purification Cylinder: made of BASF antioxidant materials and UOP water adsorbent materials
                            Moisture removal capacity: 1.3KG
                            Oxygen removal capacity: 30L
                            Stainless steel pipe connection
                            Blower: Flow Rate: 8m³/h

                            Filter System

                              1.One Organic Solvent Filter System filled with replaceable activated carbon (5.2kg)

                              2.Two pieces of HEPA filters can eliminate particle size > 0.3μm

                                Control system

                                PLC control system with 6" touchscreen display
                                Water concentration: 0 - 1000 ppm @0.1 ppm 
                                Oxygen concentration: 0 - 1000 ppm @0.1 ppm
                                Auto pressure control -12 to 12 mbar @0.1mbar

                                Feedthroughs

                                Can be 110VAC or 220VAC depending on the input source

                                Vacuum pump

                                Vacuum rate: 4.7 cfm
                                Lowest pressure: 0.4pa

                                Compliance

                                CE Certified

                                Warranty

                                One years limited warranty with lifetime technical support

                                Application Notes & Warnings

                                  1.The interconnections between the glove box chamber and the gas purification system must be unimpeded during the purification cycles. 
                                    2.The use of corrosive gases is prohibited as they will damage the water sensors.
                                      3.Regularly perform regeneration to maintain the optimum purification efficiency. 
                                        4.The H2O removing rate is largely dependent on the type of purging gas used. For faster chamber purging process, Nitrogen is preferred over Argon due to its lighter mass. 
                                          5.Corrosive liquid (such as LiPF6 electrolyte) must be remain sealed in a container inside the glove box. Otherwise, liquid vapour may condense and corrode the steel chamber and purification pipeline.


                                          Product Images

                                          Vacuum Glove Box

                                          Hot Tags: laboratory glove box, suppliers, customized, wholesale, for sale, 10 Channels OEM Cells Grading Sorter, Single-side Automatic Spot Welding Machine, Laboratory Small Electrolyte Filling Machine, Waste Gas Treatment Equipment, Automatic Cylindrical Cell Pack Assembly Line, Cylindrical Battery Cap Pressure Tester